THE BIRD SENTRYTM BLENDER
AN OXYGEN MONITOR, ANALYZER AND BLENDER IN ONE COMPACT, AFFORDABLE UNIT

SPECIFICATIONS

PERFORMANCE

Flow Range 20-100 LPM
Bleed Flow 10-120 LPM

CONTROLS

O2 Concentration Range 21%-100%
Adjustable High/Low O2
Concentration Alarm Range
18%-106% (±2%)
Power Switches "On"/"Off"

ALARMS / WARNINGS

Audible and Visual High/Low
vv O2 Concentration Alarm
Set by user
Audible Pressure
vv Differential Alarm
When inlet pressured differ by 20 PSI or more
Audible and Visual
vv Sensor Disconnect Alarm
Activated when sensor or cable are disconnected while power is "On"
"Lo Batt" Display When displayed, indicates that battery should be replaced

ELECTRICAL PNEUMATIC REQUIREMENTS

Power Source Standard 9 volt battery
Pneumatic Requirement 30 - 75 PSIG (2.1 - 5.3 kg/cm2) air and O2 gas sources

SENSORS

Accuracy ±2% O2
Expected Useful Life:
vvii Disposable Galvanic 438,000 O2 hours
vvii (One O2 hour is equal to the operation of the sensor in 1% oxygen for one hour.)
Sensor Response Time
(to 90% of total reading)
< 20 seconds

DIMENSIONS & WEIGHTS

    Dimensions
(HxWxD) 7.25" x 4.75" x 6.25"
(184mm x 127mm x 159mm)
    Weight
Unpackaged 4.5 lbs. (2.04 kg)

ORDERING INFORMATION

P/N: Description:
10209* Bird Sentry w/Disposable Galvanic Sensor and Cable
66030 Disposable Galvanic Cell
15177 Galvanic Cell Cable
*Includes Hoses & Air Inlet Filter/Water Trap
NOTE: All Bird MicroBlenders are available with N.I.S.T. Inlet Fittings.
NOTE: Product specifications are subject to change without notice.