TBird® Legacy

Controls
Power On / Standby
Modes Control
Assist Control
SIMV
CPAP
Flow Pattern Decelerating
Tidal Volume 50 to 2000 ml
Breath Rate 2-80 bpm
Peak Flow 10-140 lpm
 Spontaneous 180 lpm max.
Sensitivity 1 to 8 lpm, OFF
PEEP/CPAP 0 to 30 cmH2O
Pressure Support Off, 1-60 cmH2O
Pressure Control Off, 1-100 cmH2O
Inspiratory Time 0.2-10.0 secs.
Bias Flow 10-20 lpm
Sigh On / Off, 1 sigh breath every 100 breaths or 7 minutes, whichever occurs first, 1.5 x Set Tidal Volume
Manual Breath x1
Flow Cal On / Off
Pressure Relief 20-130 cmH2O
Monitors and Indicators
Patient Effort Yellow Indicator
Power Indicators
On Green Indicator
AC Green Indicator
Internal Battery Power Green, Yellow, Red Indicator
Internal Battery Charge Status Green, Yellow, Red Indicator
External Battery Power Green, Yellow Indicator
External Battery Charge Status Green, Yellow, Red Indicator
Airway Pressure Manometer -20 to 120 cmH2O
Total Breath Rate (f) 0 to 250 bpm
I:E Ratio (I:E) 99:1 to 1:99
Minute Volume (VE) 0 to 99.9 L
Peak Inspiratory Pressure (PIP) 0 to 140 cmH2O
Mean Airway Pressure (MAP) 0 to 99 cmH2O
Positive End Expiratory Pressure (PEEP) 0 to 99 cmH2O
Tidal Volume (Vte) 0 to 4000 ml
Alarms
High Pressure Limit 5 to 120 cmH2O
Low Pressure Limit Off, 2 to 60 cmH2O
Low Minute Volume 0.0 to 99.9 L
     (2 ranges) 0.1 to 99.9 L
High Breath Rate 3-150 bpm
Apnea Interval 10 to 60 seconds
Backup Breath Rate 12 bpm or set breath rate, whichever is greater
Ventilator Inoperative Red Indicator
Alarm Silence 60 second max.
Alarm Volume 65 - 85 dba
Physical Specifications
Dimensions
System
Width: 13.0 in (33.0 cm)
Depth: 14.5 in (37.0 cm)
Height: 11.0 in (28.0 cm)
Weight (approximate)
System
Unpacked 33 lbs (15 kg)
Shipping 42 lbs (19 kg)
Electrical Specifications
Input Voltage Requirements
AC
Voltage: 100-240 VAC
Frequency: 47-63 Hz
Power: 200 watts
Power Factor: .97
Input Fuse Specification Information
100-250V~:
Quantity: 2
Current Rating: 3.15 Amperes
Voltage Rating: 250 Volts
Type: 5x20 mm, IEC 127 Sheet 1
(Type F, fast acting, high-interrupting capacity)
Pneumatic Specifications
Oxygen Inlet
Oxygen Bleed in: 0-80 Ipm
Environmental Specifications
Environmental Temperature
Operating: +5 to 40°C
Storage/Shipping: -20 to 60°C
Relative Humidity
Operating: 15% to 95% relative, noncondensing
Storage/Shipping: 10% to 95% relative, noncondensing
Sound Level
<50 dBa (rms) at one meter
Shock and Vibration
Complies with the following standards:
Shock: IEC 68-2-27 (MIL-STD-810E)
Vibration: IEC 68-2-6, 68-2-34 (MIL-STD-810E)
Standards And Requirements Compliance
The Ventilator complies with the following FDA and international agency standards and requirements:
  • IEC 601-1 Medical Equipment Standard, 1988
  • IEC 601-1-2 Medical Electrical Equipment, Collateral Standard & Electromagnetic Compatibility, 1993
  • IEC601-2-12 Lung Ventilators, 1988
  • UL 2601-1 Medical Electrical Equipment, Part I: General Requirements For Safety (1994)

When packaged in its shipping container, the ventilator also complies with International Safe Transit Association requirements for packaged products weighing less than 100 pounds.

[Return]

 