V.I.P. BIRD Gold and Sterling

CONTROLS

Alarm Silence On/Off (I/O)
Alarm Volume 66-85dB at 1 meter
Apnea Switch 10 to 60 sec.
Assist Sensitivity Pressure Triggered 1 to 20 cmH20, OFF
Assist Sensitivity Flow Triggered
w/Inf. Sensor (all modes)
0.2 to 3 lpm, OFF
Assist Sensitivity Flow Triggered
w/Ped. Sensor (no TCPL)
1.0 to 5.0 lpm, OFF
Bias Flow w/Infant Flow Sensor 3lpm (except TCPL)
Bias Flow w/Pediatric Flow Sensor OFF, 5 lpm
Flow:

 

3 to 120 lpm (Volume & VAPS)
3 to 40 lpm (TCPL, Insp. Flow)
3 to 15 lpm (TCPL, Exp. Bias Flow)
Inspiratory / Expiratory Hold (Gold model only) 0 to 3 sec.
Inspiratory Pause (Gold model only) 0 to 2 sec.
Inspiratory Pressure 3 to 80 cmH20
Inspiratory Time 0.10 to 3.00 sec.
Leak Compensation ON/OFF
Mode Select
     Assist Control



     (S)IMV/CPAP/PS

TCPL, Volume,
Pressure Control (Gold Only),
VAPS (Gold Only)

TCPL, Volume,
Pressure Control (Gold Only),
VAPS (Gold Only)
O2 Concentration 21% to 100%
PEEP/CPAP 0 to 24 cmH20
Power ON/OFF
Pressure Relief Valve 0 to 130 cmH20
Pressure Support OFF, 1 to 50 cmH20
Rate 1 to 150 bpm
Rise Time (Gold Only) 1 to 7 (1 fastest, 7 slowest)
Termination Sensitivity 5 to 25% of peak flow, OFF
Tidal Volume 10ml to 1200ml
Waveform Select square, decelerating

MONITORS

Airway Pressure Manometer -20 to 140 cmH20 (minimum)
I:E Ratio (I:E) 1:0.1 to 1:60
Inspiratory Time (Ti) 0.05 to 60 sec.
Minute Volume (VE) 0.0 to 99.9 L w/flow sensor
OFF w/no flow sensor
Mean Airway Pressure (MAP) 1 to 120 cmH2
Peak Inspiratory Pressure (PIP) 1 to 130 cmH20
Positive End Expiratory Pressure (PEEP) 0 to 24 cmH2
Respiratory Rate (bpm) 0 to 250 bpm
Tidal Volume (VT) 0 to 9999 ml w/flow sensor
OFF w/no flow sensor

ALARMS / LIMITS

Blender Input Gas Alarm Air & Oxygen difference > +/-20 PSIG
High Breath Rate Alarm OFF, 0 to 300 bpm
High Pressure Alarm 3 to 120 cmH2
High / Prolonged Pressure Alarm PEEP/CPAP + 6 C
High Tidal Volume OFF, 2 ml to 2 L
Low Inlet Gas Pressure Alarm System pressure
< 22.5 PSIG or > 27.5 PSIG
Low Minute Volume Alarm OFF, 0.01 to 30 L
Low PEEP/CPAP Pressure Alarm Fixed at 3 cmH20 below set
PEEP/CPAP
Low Peak Pressure OFF, 3 to 120 cmH2
Pressure Support / VAPS Time Limit 0.10 to 3.00 sec.